Proof before claims

We’ll prove it before we claim it.

No accuracy figure appears on this site until VenTracer has been tested side by side with laboratory reference systems and the results are published, favourable or not.3 Here is the plan we will be held to.

In development · Pilot planned 2027 · Not for saleDesigned in India

Validation plan

The study, before the claims.

Example limits are shown to make the plan concrete; final limits will be fixed and registered in the protocol before data collection. How the methods and checks feed into it

ElementPlanStatus
Before testing
ReferenceA laboratory metabolic cart, tested on the same peoplePlanned
BenchMetabolic simulator across 0.5–5 L/min V̇O₂. V̇E 20–100 L/min with the current flow element; 20–200 L/min with the high-flow element once built. Each element reported separately, including where it stops being accuratePlanned
Pre-registrationProtocol, acceptance limits and analysis plan registered publicly (for example on OSF) before the first participant is tested. Any deviation reported in the paperPlanned
Acceptance limitsSet before testing and justified from published reference-system agreement and the smallest worthwhile change in VO₂max, not chosen after seeing data (for example V̇O₂ bias within ±3% and 95% limits of agreement within ±7%; VT power within ±15 W)Planned
Sample and populationSample size justified in advance from the expected limits of agreement. Participants span VO₂max about 25–75 ml/kg/min, both sexes and ages 18–70Planned
Frozen versionsFirmware, algorithm and app versions frozen and named for the whole study; results apply only to those versionsPlanned
During testing
ProtocolSeated rest, then an incremental ramp to exhaustionPlanned
ConditionsCycling and treadmill running; a warm and humid room. V̇E above 150 L/min is tested only with the high-flow element, once it exists. Each flow element is validated and reported separatelyPlanned
Reference quality controlThe reference cart is calibrated to the manufacturer’s procedure before each session and checked against a metabolic simulator or with Douglas bags. Its own error is reportedPlanned
Measurement configurationWhere possible, both systems measure the same breaths (in series, with added dead space and resistance reported). Otherwise, repeated randomised-order tests. The configuration is stated in the paperPlanned
ReliabilityRepeat tests on separate days, reported as typical error and smallest worthwhile changePlanned
After testing
StatisticsBland–Altman bias and 95% limits of agreement (absolute and %), a test for proportional bias, and equivalence testing against the pre-registered limits. Correlation and ICC are reported but not used to judge agreementPlanned
ThresholdsCompared three ways: VenTracer algorithm vs cart (expert pick), expert picks on VenTracer data vs expert picks on cart data, and VenTracer algorithm vs experts on VenTracer data. Inter-rater agreement reportedPlanned
Complete reportingAll tests reported, including failed tests, leak-flagged breaths, out-of-range breaths and dropouts. No post-hoc exclusionsPlanned
IndependencePartner labs analyse their own data; the paper is submitted to a peer-reviewed journal whatever the resultPlanned
PublicationFull results published, favourable or not, with de-identified data and analysis code, before any accuracy figure appears on this siteCommitted
Published results

None yet.

VenTracer has no published validation data. Until it does, readings from early units are experimental.1 When results exist, they will be listed here with the full paper, data and methods. The agreement chart for them is drawn on the methods page, and it stays empty until the data exist.

The empty agreement chart

What VenTracer can’t do (yet)

  • Claim accuracyNot until validation is published
  • Cover maximal tests in trained athletesThe current flow element is characterised to about 85–100 L/min; a high-flow element is in development
  • Find thresholds from short or broken testsA proper ramp is needed for a clear VT2
  • Work through a leaking maskA seal check before every test and a leak check on every breath are planned
  • Diagnose anythingIt is not a medical device4

All known limitations

Roadmap

From concept to published data.

  1. Now · 2026

    Concept and architecture

    System design, sensing plan, industrial design and brand.

  2. 2027 · planned

    Engineering prototypes

    First working analyzers on the bench and on athletes.

  3. 2027 · planned

    Validation pilot

    Units tested side by side with partner labs’ and clinics’ reference systems.

  4. After validation

    Wider pilot

    Units for coaches, athletes and clinics, once results are published.

Timelines are plans, not commitments, and may change with development and validation results.